MICROLITER FLUID CALIBRATOR™
PPM CALIBRATION SAMPLES OF MOISTURE IN PROCESS GAS OR LIQUID STREAMS

Model 203P (96k)

Calibrator for Moisture Analyzers
Model #203P

FEATURES:

VERSATILE Injects liquid or gas dissolved in liquid to calibrate analyzers in PPM ranges
BROAD RANGING Prepares calibration mixtures in a flowing stream of a few PPM to percent or meter by timing into a fixed volume
EXPLOSION PROOF Equipment housed in aluminum enclosure rated for Class 1, Group B or C & D
AUTOMATIC CALIBRATION Optional timer control provides unattended calibration
RATE ADJUSTABLE Injection rate can be easily adjusted for wide PPM ranging of carcinogens
PULSE RATE METER A six digit meter allows reproducible inject rates from 6 to 1,000 steps per minute
TEMPERATURE No thermostat controls required
HIGH PRESSURE Injection can be into the sample stream under high pressure
MIXTURES Mix 2 or more miscible liquids for several standards


CALIBRATE MOISTURE ANALYZERS

The  Model 203 "Microliter Fluid Calibrator™" will accept almost any fluid needed for injection into a flowing stream. Minute quantities of fluid are rapidly "pulsed" into a stream to effectively produce a continuous flowing mixture suitable for calibration purposes. Very low PPM to per cent levels of fluid in the stream are easily produced with this model. Equally well, "tracer compounds" or "chemical additives" can be blended into a gas or liquid stream under process conditions knowing the flow rate of the sample stream.

A control is provided for adjusting the instrument's pulse rate to achieve a desired reading on the digital meter display. This adjustment establishes a precise drive rate. The rate chosen governs the speed of travel of a piston seal within a fluid reservoir. The combination of drive rate and reservoir capacity fixes the rate of fluid leaving the reservoir for entry into flowing stream. Knowing this fluid flow rate and the flow rate take out of the stream, the concentration in PPM of the fluid in the stream is known.

A 3-way valve is provided to permit fluid to be charged to the reservoir conveniently. Optional solenoid and calibration timer for automatic operation are available.

| Home |
| Permeation Trans. Monitor | Volatile Hydro. Anal. |
| Membrane Sample Trans. | Membrane Probe Trans. |
| Permeation Tube Calibrator | Permeation Tubes |
| Microliter Fluid Calibrator | Sensor Calibration |
| Membrane Transfer Technology | TNRCC Letter |

Phone: 281-516-3950 Fax: 281-351-8925
E-Mail: Sales@AsiWebPage.com
Keco R&D Inc.

™ Trademark Keco R.&D., Inc.
Patent # 5,156,334 and Patent Pending
© 1994-2001 Keco R.& D., Inc.